DIGICARB 3:Thermal analyser
DIGICARB 3 is a microprocessor based digital instrument for the automatic thermal analysis of molten iron. It is designed for use with plug - in, fast response, Type 'K' Thermocouple sample cups.

The instrument is housed in a steel, dust - proof case which may be bench, panel or wall mounted.
 
digicar 3 thermal analyser,

AUTOMATIC FEATURES :

  • Cooling Curve validity check
  • Calculation of % C.E.L. (or C.E.V.), % C, % Si and delta T
  • Curve progress indication
  • Plateau detection
  • Metal temperature measurement

The instrument is programmed to check validity of the cooling curve and will reject poor readings, indicating the reason for rejection.
A numbering system indicates the State of progress through the cooling curve.
An integral printer allows a permanent record of each set of requirements plus time, date and %P.
The unique trend program enables a listing of the previous 40 sets of analysis to be printed.

An RS232/485 Serial Output port gives a download facility to a remote printer or computer, for storage/processing of results. A Windows 95/98 Download Disk is provided.

SPECIFICATION :

  • RANGE :
    0 to 1372 deg. C for NiCr/Ni/Al Type 'K' Thermocouple (0 to 2500 deg. F)
    2.50 to 4.80 % for C.E.L. or C.E.V. %
    2.10 to 4.20 % Carbon
    0.50 to 3.60 % Silicon
    0.00 to 0.99 % Phospherous
    -0.99 to +0.99 % Si Offset
  • RESOLUTION :
    1 deg. C or 1 deg. F for temperatures
    0.01 % for all other measurements
  • ACCURACY :
    Temperature : +/- 0.1 % of reading +/- 1 deg. C or F
    Analysis : +/- 0.1%
  • OPERATING RANGE :
    0 to 45 deg. C (32 to 113 deg. F)
  • AUTOMATIC THERMOCOUPLE COLD JUNCTION COMPENSATION
  • DIMENSIONS : 600x335x320 mm
  • WEIGHT : 33kg
  • POWER SUPPLY :
    100, 110, 220 or 230V, 50/60 Hz