An 8 positon switch allows selection for continous monitoring
of :
- Sample
Temperature
- Liquidus
Temperature
- Solidus
Temperature
- Carbon
Equivalent Liquidus or C. E.Value %
- Carbon
%
- Silicon
%
- Phosphorous
% - a present, adjustable Value
- Silicon
Offset - a present, adjustable Value
Silicon
Offset adjustment allows the effect of residual elements on
the Arrest Temperatures, to be cancelled out. Alternatively,
it may be used to enable the Si reading to be tuned to the
Spectrum Analyser value.
Inside
Panel switches enable selection off :
- C.E.L.
or C.E.V. %
- Degrees
Celsius or Farenheit
- BCIRA
for Tectip/Carbon Cup or Quick - cup formula
- Measure
or Calibrate mode, allows instrument to be used adittionally,
as a Carbon and Silicon calculator.
- UP/DOWN
(Increase/Decrease) buttons, and in Calibrate mode, enable
P %, Si Offset % and other parameters to be adjusted : DIGICARB
2D can be fine tuned to your Cooling Curve.
An
RS232/485 Serial Output port gives a download facility to
a remote printer or computer, for storage/processing of results.
A Windows 95/98 Download Disk is provided.
SPECIFICATION
:
- RANGE
:
0 to 1372 deg. C for NiCr/Ni/Al Type 'K' Thermocouple (0
to 2500 deg. F)
2.50 to 4.80 % for C.E.L. or C.E.V. %
2.10 to 4.20 % Carbon
0.50 to 3.60 % Silicon
0.00 to 0.99 % Phospherous
-0.99 to +0.99 % Si Offset
- RESOLUTION
:
1 deg. C or 1 deg. F for temperatures
0.01 % for all other measurements
- ACCURACY
:
Temperature : +/- 0.1 % of reading +/- 1 deg. C or F